Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Declared Accuracy*: The table below shows Arms values measured using the Model 3230 in a
clinical study.

Range Specified Oxygen Saturation (Arms) Finger Oxygen Saturation (Arms) Thumb Oxygen Saturation (Arms) Low Perfusion Oxygen Saturation (Arms)
70 – 100% ±2 ± 1.31 ± 1.56 ±2
70 – 80% ±2 ± 1.65 ± 1.91 ±2
80 – 90% ±2 ± 1.05 ± 1.21 ±2
90 – 100% ±2 ± 1.18 ± 1.49 ±2


This graph shows plots of the
error (SpO2 – SaO2) by SaO2
using the 3230 with a linear
regression fit and upper 95% and
lower 95% limits of agreement.
Each sample data point is
identified by subject from a
clinical study in non-motion
conditions.

Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red. 660 nanometers @ 0.8 mW max. average
Infrared. 910 nanometers @ 1.2 mW max. average

Temperature:
Operating. -5 °C to 40 °C / 23 °F to 104 °F
Storage/Transportation. -40 °C to 70 °C / -40 °F to 158 °F

Humidity:
Operating. 10% to 95% non-condensing
Storage/Transportation. 10% to 95% non-condensing

Altitude:
Operating. Up to 4,000 meters / 13,123 feet
Hyperbaric Pressure. Up to 4 atmospheres

Battery Life:
Operating. Approximately 2,200 spot checks (25 sec. per spot-check), within 10
meters/32 feet of collector with streaming data
Storage. 1 month, with batteries installed. CAUTION: Remove batteries if the
device will be stored for more than 30 days.

* ±1 Arms represents approximately 68% of measurements.
** This information is especially useful for clinicians performing photodynamic therapy.

Model 3230 Bluetooth LE Troubleshooting

Frequently Asked Questions

Nonin
Locations

Nonin Medical Inc
(Global Headquarters)

Address
13700 1st Ave N,
Plymouth, MN 55441

Email
info@nonin.com

Toll Free
1.800.356.8874

Phone
+1.763.553.9968

Nonin Medical B.V.
(Europe)

Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
The Netherlands

Email
infointl@nonin.com

Toll Free
+31 (0)13-79 99 040

Phone
+31 (0)13-79 99 042

Interested in a career
with Nonin?