3500SP

Resources

3500SP Specifications

Print Specs

Displayed Carboxyhemoglobin Range (COHb): 0 to 40%
Displayed Met Hemoglobin Range (MetHb): 0 to 40%
Displayed Oxygen Saturation Range (SpO2): 0 to 100%
Displayed Pulse Rate Range: 18 to 321 beats per minute (BPM)
Displayed Regional Saturation Range (rSO2): 50% to 100%
SpO2 Measurement Wavelengths and Output Power*: Two wavelengths between 660 and 910 nanometers with output power range between 0.8 – 3.0 mW nominal.
CO-Met Measurement Wavelengths and Output Power*: Eight wavelengths between 600 and 920 nanometers with output power range between 1.5 and 18 mW
rSO2 Measurement Wavelengths and Output Power*: Three or four wavelengths between 730 and 880 nanometers with output power range between 3.0 and 4.5 mW max. avg.

SpO2 Accuracy (Arms**): 70 to 100%

No Motion Adults/Pediatrics*** Neonates
REUSABLE:
8100AX Series: ± 2 digits N/A
8100SX Series: ± 2 digits N/A
8100Q2: ± 2 digits N/A
DISPOSABLE:
6100CX Series: ± 2 digits ± 3 digits
Motion
REUSABLE:
8100AA: ± 2 digits N/A
8100AP: ± 3 digits N/A
8100SX Series: ± 3 digits N/A
DISPOSABLE:
6100CA/6100CP/6100CI ± 2 digits N/A
Low Perfusion****
8100AX, 8100Q2, 6100C(X) 70 – 97% ± 2 digits N/A
8100S(X) 70 – 100% ± 2 digits N/A
Pulse Rate Accuracy Adults/Pediatrics*** Neonates
No Motion
REUSABLE:
8100AX Series: 40 – 240 BPM ± 3 digits N/A
8100SX Series: 18 – 300 BPM ± 3 digits N/A
8100Q2: 40 – 240 BPM ± 3 digits N/A
DISPOSABLE:
6100CX Series: 20 – 250 BPM ± 3 digits 20 – 250 BPM ± 3 digits
Motion (40 – 240 BPM)
REUSABLE:
8100AX Series: ± 5 digits N/A
8100SX Series: ± 5 digits N/A
Low Perfusion (40 – 240 BPM) **** ± 3 digits ± 3 digits

* This information is especially useful for clinicians performing photodynamic therapy.
** ±1 Arms represents approximately 68% of measurements at zero bias.
*** Includes infant patients.
**** Does not apply to those sensors listed as N/A under the neonate column and 8100Q2.

Notes:
SpO2 Reusable Group:
8100AX Finger Clip Sensors: 8100AA-1, 8100AP-1
8100SX Soft Sensors: 8100SS, 8100SM, 8100SL,
8100Q2 Ear Clip Sensor

SpO2 Disposable Group:
6100C Series Sensors: 6100CA, 6100CP, 6100CI, 6100CN
CO-Met Reusable Group: 8300AA
rSO2 Reusable Group: 8003CA, 8004CA, 8004CB, 8004CB-NA
rSO2 Disposable Group: 8204CA

8300AA CO-Met Accuracy:

Description (No Motion) SaO2 Range MetHb or COHb Range Accuracy
%SpO2 accuracy with COHb 70 – 100% 0 – 3% COHb ± 2
%SpO2 accuracy with COHb 80 – 100% 0 – 15% COHb ± 2
%SpO2 accuracy with MetHb 70 – 100% 0 – 2% MetHb ± 2
%SpO2 accuracy with MetHb 80 – 100% 0 – 15% MetHb ± 2
%COHb accuracy 95 – 100% 0 – 15% COHb ± 2
%COHb accuracy with elevated HHb 80 – 100% 0 – 15% COHb ± 3
%MetHb accuracy 97 – 100% 0 – 15% MetHb ± 0.8
%MetHb accuracy with elevated HHb 80 – 100% 0 – 15% MetHb ± 1.0

rSO2 Accuracy

8004CA, 8204CA: 50 – 100%

Accuracy Right Left Both Hypercapnia Hypocapnia
Absolute 4.1 3.8 3.9 5.1 3.3
Trend 1.9 3.0 2.5 3.4 3.8

8004CB, 8004CB-NA: 45 – 95% ±5.9
8003CA: 45 – 100%

Accuracy Right Left Both Hypercapnia Hypocapnia
Trend 2.3 4.6 3.6 6.6 3.7

Temperature (Operating):
Signal Processor: -5 to +50 °C
8100AX Series, 8100Q2, 8300AA: -10 to +40 °C
8100SX Series, 6100C(X) Series: 0 to +40 °C
rSO2 sensors: -5 to +40 °C

Storage/Transportation:
Signal Processor, 8100SX Series, 6100C(X) Series, rSO2 sensors: -30 to +70 °C
8100AX Series, 8100Q2, 8300AA: -40 to +70 °C

Humidity (Operating):
SpO2 and CO-Met sensors: 15% to 93% non-condensing
rSO2 sensors: 10% to 90% non-condensing

Storage/Transportation:
SpO2 and CO-Met sensors: Up to 93% non-condensing
rSO2 sensors: 10% to 95% non-condensing

Altitude (Operating): 0 to 4,000 meters (13,123 feet)
Patient Isolation: Meets IEC 60601-1 Dielectric withstand
Degree of protection: Type BF-Applied Part

Leakage Current: Not Applicable
Dimensions (approximate):
Cable Length: .75 meter ±10%
Housing: 76.2 mm (3 in.) x 21.6 mm (.85 in.) x 25.4 mm (1 in.)

Weight (approximate): 50 grams
Power Draw (typical):
No Sensor Attached: 210 mW
SpO2 Sensor Attached: 310 mW
rSO2 Sensor Attached: 315 mW
CO-Met Sensor Attached: 360 mW

Maximum Power Draw:
No Sensor Attached: 280 mW
SpO2 Sensor Attached: 370 mW
rSO2 Sensor Attached: 385 mW
CO-Met Sensor Attached: 490 mW

Inrush Power1: An average of 810 mW within a sliding 5 mS window for any voltage input from 3.82 to 5.40 VDC.
Voltage Input: t5 ± 0.25 VDC, w/100 mV max. Ripple
Minimum input voltage at signal processor connector: 3.82 VDC
Maximum input voltage at signal processor connector: 5.4 VDC
Minimum slope: 10 V/s (i.e. Maximum rise time 0.5 seconds)

Data I/O Signals: RS-485 type UBUS: S- line must be at least 200 mV lower than the S+ line when driven
Fluid Spill Resistance: IP22
Ruggedness: IEC 60068-2-27
Shock: Sinusoidal – IEC 60068-2-6
Ruggedness (Vibration): Random – IEC 60068-2-64

Sensors: Designed to use Nonin branded CO-Met sensors and Nonin-branded PureLight® SenSmart sensors (see Accessories)
Shielding: An RF shield is included (placed over the analog components)
Regulatory Status/Information: Contact Nonin regulatory@nonin.com
ROHS: 2011/65/EU
Biocompatability: ISO 10993-01

This product complies with all applicable clauses of the following standards:
IEC 60601-1, 3rd Ed
IEC 60601-1-11:2010
ISO 9919:2005
ISO 80601-2-61

1 Dependent upon type of sensor connected.
t Nonin recommendation for input voltage. Voltage drops across extension cables will reduce the amount of voltage delivered to the device and must be taken into account.

3500SP Troubleshooting

Frequently Asked Questions

Nonin
Locations

Nonin Medical Inc
(Global Headquarters)

Address
13700 1st Ave N,
Plymouth, MN 55441

Email
info@nonin.com

Toll Free
1.800.356.8874

Phone
+1.763.553.9968

Nonin Medical B.V.
(Europe)

Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
The Netherlands

Email
infointl@nonin.com

Toll Free
+31 (0)13-79 99 040

Phone
+31 (0)13-79 99 042

Interested in a career
with Nonin?