6500 Durafoam Specifications
SpO2 Accuracy:1, 2, 3
Range | Oxygen Saturation (Arms*) |
---|---|
70 – 100% | ±2 |
70 – 80% | ±3 |
80 – 90% | ±2 |
90 – 100% | ±2 |
SpO2 Low Perfusion Accuracy: 70% to 100% ±2 digits (Arms*)1
Pulse Rate Accuracy: 18 BPM to 300 BPM ±3 digits (Arms*)1
Pulse Rate Low Perfusion Accuracy: 40 BPM to 240 BPM ±3 digits (Arms*)1
Temperature: 4, 5
Operating: -20 °C to 50 °C (-4 °F to 122 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)
Humidity: 4, 5
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
* ±1 Arms encompasses 68% of the population at zero bias.
1 Additional accuracy and performance information can be found in the sensor accuracy document on the operator’s manual CD.
2 Accuracy testing was performed under no-motion conditions.
3 Accuracy specifications based on Nonin’s PureSAT® SpO2 technology and PureLight® sensor technology.
4 For combined oximeter/sensor specifications, refer to the applicable oximetry system’s operator’s manual.
5 Range as tested with Nonin’s PureSAT SpO2 technology.
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW nominal
Infrared: 910 nanometers @ 1.2 mW nominal
** This information is especially useful for clinicians performing photodynamic therapy.
Compliance
This product complies with ISO 10993-1.
Not made from natural rubber latex.
Nonin
Locations
Nonin Medical Inc
(Global Headquarters)
Address
13700 1st Ave N,
Plymouth, MN 55441
Email
info@nonin.com
Toll Free (USA and Canada)
1.800.356.8874
Phone
1.763.553.9968
Nonin Medical B.V.
(Europe)
Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
Netherlands
Email
info@nonin.com