Model 8000A Series Reusable Finger Clip Pulse Oximetry Sensors Specifications
SpO2 Accuracy1, 2:
Range | Oxygen Saturation (Arms*) (figure A) | Motion Oxygen Saturation(Arms*) (figure B) |
---|---|---|
70 – 100% | ±2 | ±3 |
70 – 80% | ±2 | ±3 |
80 – 90% | ±2 | ±3 |
90 – 100% | ±2 | ±2 |
A
B
8000AA Sensors
SpO2 Low Perfusion Accuracy: 70 to 100% ±2 digits (Arms*)1
Pulse Rate Accuracy: 18 to 300 BPM ±3 digits (Arms*)1
Pulse Rate Low Perfusion Accuracy: 40 to 240 BPM ±3 digits (Arms*)1
Temperature: 3, 4
Operating: -20 °C to 50 °C (-4 °F to 122 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)
8000AP Sensors
SpO2 Low Perfusion Accuracy: 70 to 100% ±2 digits (Arms*)1
Pulse Rate Accuracy: 40 to 240 BPM ±3 digits (Arms*)1
Pulse Rate Motion Accuracy: 40 to 250 BPM ±5 digits (Arms*)1
Pulse Rate Low Perfusion Accuracy: 40 to 240 BPM ±3 digits (Arms*)1
Temperature: 3, 4
Operating: -20 °C to 50 °C (-4 °F to 122 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)
Humidity: 3, 4
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
* ±1 Arms encompasses 68% of the population at zero bias.
1 Additional accuracy and performance information can be found in the sensor accuracy document on the operator’s manual CD.
2 Accuracy specifications based on Nonin’s PureSAT® SpO2 technology and PureLight® sensor technology.
3 For combined oximeter/sensor specifications, refer to the applicable oximetry system’s operator’s manual.
4 Range as tested with Nonin’s PureSAT SpO2 technology.
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 3 mW nominal
Infrared: 910 nanometers @ 3 mW nominal
** This information is especially useful for clinicians performing photodynamic therapy.
Compliance
This product complies with ISO 10993.
Not made from natural rubber latex.