Model 8000A Series Reusable Finger Clip Pulse Oximetry Sensors

Model 8000A Series Reusable Finger Clip Pulse Oximetry Sensors Specifications

Print Specs

SpO2 Accuracy1, 2:

Range Oxygen Saturation (Arms*) (figure A) Motion Oxygen Saturation(Arms*) (figure B)
70 – 100% ±2 ±3
70 – 80% ±2 ±3
80 – 90% ±2 ±3
90 – 100% ±2 ±2

A

B

SpO2 Low Perfusion Accuracy: 70 to 100% ±2 digits (Arms*)1
Pulse Rate Accuracy: 40 to 240 BPM ±3 digits (Arms*)1
Pulse Rate Motion Accuracy: 40 to 250 BPM ±5 digits (Arms*)1
Pulse Rate Low Perfusion Accuracy: 40 to 240 BPM ±3 digits (Arms*)1
Temperature: 3, 4
Operating: -20 °C to 50 °C (-4 °F to 122 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)

Humidity: 3, 4
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing

* ±1 Arms encompasses 68% of the population at zero bias.
1 Additional accuracy and performance information can be found in the sensor accuracy document on the operator’s manual CD.
2 Accuracy specifications based on Nonin’s PureSAT® SpO2 technology and PureLight® sensor technology.
3 For combined oximeter/sensor specifications, refer to the applicable oximetry system’s operator’s manual.
4 Range as tested with Nonin’s PureSAT SpO2 technology.

Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 3 mW nominal
Infrared: 910 nanometers @ 3 mW nominal
** This information is especially useful for clinicians performing photodynamic therapy.

Compliance
This product complies with ISO 10993.
Not made from natural rubber latex.

Model 8000A Series Reusable Finger Clip Pulse Oximetry Sensors Troubleshooting

Frequently Asked Questions

Nonin
Locations

Nonin Medical Inc
(Global Headquarters)

Address
13700 1st Ave N,
Plymouth, MN 55441

Email
info@nonin.com

Toll Free
1.800.356.8874

Phone
+1.763.553.9968

Nonin Medical B.V.
(Europe)

Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
The Netherlands

Email
infointl@nonin.com

Toll Free
+31 (0)13-79 99 040

Phone
+31 (0)13-79 99 042

Interested in a career
with Nonin?