SenSmart Multi-Sensing 8330AA Specifications
Overall Accuracy:
Accuracy (Arms) is expected to encompass approximately 68% of measured values and 95% Limits of Agreement are expected to encompass approximately 95% of measured values.
Description SaO2 Range MetHb or COHb Range Clinical Accuracy Expected Accuracy/95% Limits of Agreement
%SpO2 accuracy with COHb 70 - 100% 0 - 3% COHb ± 2 (-2.6, 3.2)
%SpO2 accuracy with COHb 80 - 100% 0 - 15% COHb ± 2 (-2.3, 2.9)
%SpO2 accuracy with MetHb 70 - 100% 0 - 2% MetHb ± 2 (-2.3, 3.7)
%SpO2 accuracy with MetHb 80 - 100% 0 - 15% MetHb ± 2 (-3.0, 3.8)
%COHb accuracy with elevated HHb 95 - 100% 0 - 15% COHb ± 3 (-5.2, 4.3)
%COHb accuracy with elevated HHb 80 - 100% 0 - 15% COHb ± 3 (-5.5, 4.5)
%MetHb accuracy 97 - 100% 0 - 15% MetHb ± 1 (-1.5, 1.7)
%MetHb accuracy with elevated HHb 80 - 100% 0 - 15% MetHb ± 1 (-1.7, 2.1)
Observed Clinical Study Accuracy
The tables below show Arms values measured using the Model 8330AA in a clinical study of healthy volunteer subjects.
%SpO2 Clinical Accuracy (Arms)1,2:
SaO2 Range | COHb Range | Oxygen Saturation (Figure A) |
---|---|---|
70 - 100% | 0 - 3% | ± 2 |
70 - 80% | 0 - 3% | ± 2 |
80 - 90% | 0 - 3% | ± 3 |
90 - 100% | 0 - 3% | ± 2 |
SaO2 Range | COHb Range | Oxygen Saturation with COHb (Figure B) |
---|---|---|
80 - 100% | 0 - 15% | ± 2 |
80 - 90% | 0 - 15% | ± 3 |
90 - 100% | 0 - 15% | ± 2 |
SaO2 Range | MetHb Range | Oxygen Saturation (Figure C) |
---|---|---|
70 - 100% | 0 - 2% | ± 2 |
70 - 80% | 0 - 2% | ± 2 |
80 - 90% | 0 - 2% | ± 2 |
90 - 100% | 0 - 2% | ± 2 |
SaO2 Range | MetHb Range | Oxygen Saturation with MetHb (Figure D) |
---|---|---|
80 - 100% | 0 - 15% | ± 2 |
80 - 90% | 0 - 15% | ± 3 |
90 - 100% | 0 - 15% | ± 2 |
%COHb Accuracy (Arms):
SaO2 Range | COHb Range | %COHb Accuracy (Figure E) |
---|---|---|
95 - 100% | 0 - 15% | ± 3 |
95 - 100% | 0 - 5% | ± 2 |
95 - 100% | 5 - 10% | ± 3 |
95 - 100% | 10 - 15% | ± 3 |
SaO2 Range | COHb Range | %COHb Accuracy with Elevated HHb (Figure F) |
---|---|---|
80 - 100% | 0 - 15% | ± 3 |
80 - 100% | 0 - 5% | ± 3 |
80 - 100% | 5 - 10% | ± 3 |
80 - 100% | 10 - 15% | ± 3 |
%MetHb Accuracy (Arms):
SaO2 Range | MetHb Range | %MetHb Accuracy (Figure G) |
---|---|---|
97 - 100% | 0 - 15% | ± 1 |
97 - 100% | 0 - 5% | ± 1 |
97 - 100% | 5 - 10% | ± 1 |
97 - 100% | 10 - 15% | ± 1 |
SaO2 Range | MetHb Range | %MetHb Accuracy with Elevated HHb (Figure H) |
---|---|---|
80 - 100% | 0 - 15% | ± 1 |
80 - 100% | 0 - 5% | ± 1 |
80 - 100% | 5 - 10% | ± 1 |
80 - 100% | 10 - 15% | ± 1 |
Pulse Rate Accuracy: 40 – 250 BPM ±3 digits (Arms):
Temperature:
Operating: -10 °C to 40 °C (14 to 104 °F)
Transient Operating: -20 °C to 50 °C (-4 to 122 °F)
Storage/Transportation: -40 °C to 70 °C (-40 to 158 °F)
Humidity:
Operating: 15% to 93% non-condensing
Transient Operating: 15% to 90% non-condensing
Storage/Transportation: Up to 93% non-condensing
Operating Altitude: 0 to 4,000 meters (0 to 13,123 feet)
Measurement Wavelengths and Output Power
Wavelength range: 600 to 910 nanometers
Output range: 1.5 to 18 mW
Compliance
This product complies with ISO 10993-1.
Not made with natural rubber latex.
Warranty
For warranty information refer to: https://www.nonin.com/warranty/
Nonin
Locations
Nonin Medical Inc
(Global Headquarters)
Address
13700 1st Ave N,
Plymouth, MN 55441
Email
info@nonin.com
Toll Free (USA and Canada)
1.800.356.8874
Phone
1.763.553.9968
Nonin Medical B.V.
(Europe)
Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
Netherlands
Email
info@nonin.com