Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 300 pulses per minute (BPM)
Displays:
Numeric Displays: 3-digit LEDs, Tricolor (red, green, amber)
Pulse Strength Bar Graph LED: Tri-color LED segments
Accuracy – Sensors: Declared accuracy data for compatible sensors can be found in Nonin’s Sensor Accuracy document.
Measurement Wavelengths and Output Power*:
Red: 660 nanometers @ 0.8 mW maximum average
Infrared: 910 nanometers @ 1.2 mW maximum average
Alarm Volume Range: 52–82 dBA
Informational Tone Volume Range: 50–70 dBA
Temperature:
Operating: 32 °F to 122 °F (0 °C to 50 °C)
Storage/Transportation: -40 °F to 158 °F (-40 °C to 70 °C)
Humidity:
Operating: 10% to 90% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude (Operating): Up to 12,000 meters (40,000 feet)
Power Requirements: 7.2 volt battery pack (6 cells) or 12 VDC 1.5A AC power supply
Battery Life:
Operating: Minimum 12 hours of continuous operation with a fully charged battery pack
Storage: 27 days
Battery Recharge Time: 4 hours
Dimensions: 5.5” H x 7.25” W x 4.5” D
Weight: 2.2 lbs
Memory: 115 hours minimum
* This information is especially useful for clinicians performing photodynamic therapy.
Classifications per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1/ UL60601-1:
Type of Protection: Class II (on AC power with MPP30 power supply). Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Enclosure Degree of Ingress Protection: IP22
Mode of Operation: Continuous