Model 9847V

Model 9847V Specifications

Print Specs

Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 450 beats per minute (BPM)
CO2 Range: 0 to >75mmHg
CO2 Response Time: 250 ms
Respiration Rate Range: 1 to 60 breaths per minute
Breath Detection Threshold: 5 mmHg
Saturation Declared Accuracy (Arms):* 70 – 100% ±3 digits for the Model 2000SL, 2000SA, and 2000T sensors.
Below 70% is not specified for all sensors.

Pulse Rate Declared Accuracy (Arms)*: ± 3% ± 1 digit
CO2 Accuracy of Bar Graph Thresholds (Arms)*: ±25% of reading (typical)
Measurement Wavelengths and Output Power**::
Red: 660 nanometers @ 0.8 mW maximum average
Infrared: 910 nanometers @ 1.2 mW maximum average

Alarm Volume Range: 70 – 71 dBA
Indicators:
Pulse Quality Indicator: LED, tricolor
Numeric Displays: 3-digit, 7-segment LEDs, tricolor
CO2 Bar Graph: 8-segment bar graph, red
Low Battery Indicator: Dedicated icon, yellow
No Breath Indicator: Dedicated icon, red
Audible Alarm Disabled Indicator: Dedicated icon, yellow
Audible Indicator: Miniature speaker

Temperature:
Operating: 0 °C to +40 °C (32 °F to +104 °F)
Temperature (Storage/Transportation): -40 °C to +70 °C (-40 °F to +158 °F)

Humidity:
Operating: 10% to 95% noncondensing
Storage/Transportation: 10% to 95% noncondensing

Altitude:
Operating: Up to 3,000 meters (10,000 feet)
Hyperbaric Pressure: Up to 4 atmospheres

Power Requirements:
Six 1.5V AA-size alkaline batteries.
90 hours – Pulse Oximeter only.
20 hours – CO2 and Pulse Oximeter.
24 hours – CO2 only.

Dimensions: 8 cm W x 15 cm H x 2.5 cm D (3 in W x 6 in H x 1 in D)
Weight: 310 g (11 oz) (with alkaline batteries)
Classifications per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1 / UL 60601-1:
Type of Protection: Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous

Enclosure Degree of Ingress Protection: IP32

This device is not made with natural rubber latex.

* ± 1 Arms represents approximately 68% of measurements.
** This information is especially useful for clinicians performing photodynamic therapy.

Model 9847V Troubleshooting

Frequently Asked Questions

Nonin
Locations

Nonin Medical Inc
(Global Headquarters)

Address
13700 1st Ave N,
Plymouth, MN 55441

Email
info@nonin.com

Toll Free
1.800.356.8874

Phone
+1.763.553.9968

Nonin Medical B.V.
(Europe)

Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
The Netherlands

Email
infointl@nonin.com

Toll Free
+31 (0)13-79 99 040

Phone
+31 (0)13-79 99 042

Interested in a career
with Nonin?