Model 9847V

Model 9847V Specifications

Print Specs

Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 450 beats per minute (BPM)
CO2 Range: 0 to >75mmHg
CO2 Response Time: 250 ms
Respiration Rate Range: 1 to 60 breaths per minute
Breath Detection Threshold: 5 mmHg
Saturation Declared Accuracy (Arms):* 70 – 100% ±3 digits for the Model 2000SL, 2000SA, and 2000T sensors.
Below 70% is not specified for all sensors.

Pulse Rate Declared Accuracy (Arms)*: ± 3% ± 1 digit
CO2 Accuracy of Bar Graph Thresholds (Arms)*: ±25% of reading (typical)
Measurement Wavelengths and Output Power**::
Red: 660 nanometers @ 0.8 mW maximum average
Infrared: 910 nanometers @ 1.2 mW maximum average

Alarm Volume Range: 70 – 71 dBA
Indicators:
Pulse Quality Indicator: LED, tricolor
Numeric Displays: 3-digit, 7-segment LEDs, tricolor
CO2 Bar Graph: 8-segment bar graph, red
Low Battery Indicator: Dedicated icon, yellow
No Breath Indicator: Dedicated icon, red
Audible Alarm Disabled Indicator: Dedicated icon, yellow
Audible Indicator: Miniature speaker

Temperature:
Operating: 0 °C to +40 °C (32 °F to +104 °F)
Temperature (Storage/Transportation): -40 °C to +70 °C (-40 °F to +158 °F)

Humidity:
Operating: 10% to 95% noncondensing
Storage/Transportation: 10% to 95% noncondensing

Altitude:
Operating: Up to 3,000 meters (10,000 feet)
Hyperbaric Pressure: Up to 4 atmospheres

Power Requirements:
Six 1.5V AA-size alkaline batteries.
90 hours – Pulse Oximeter only.
20 hours – CO2 and Pulse Oximeter.
24 hours – CO2 only.

Dimensions: 8 cm W x 15 cm H x 2.5 cm D (3 in W x 6 in H x 1 in D)
Weight: 310 g (11 oz) (with alkaline batteries)
Classifications per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1 / UL 60601-1:
Type of Protection: Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous

Enclosure Degree of Ingress Protection: IP32

This device is not made with natural rubber latex.

* ± 1 Arms represents approximately 68% of measurements.
** This information is especially useful for clinicians performing photodynamic therapy.

Nonin
Locations

Nonin Medical Inc
(Global Headquarters)

Address
13700 1st Ave N,
Plymouth, MN 55441

Email
info@nonin.com

Toll Free (USA and Canada)
1.800.356.8874

Phone
1.763.553.9968

Nonin Medical B.V.
(Europe)

Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
Netherlands

Email
info@nonin.com



Interested in a career
with Nonin?