OEM III Module

OEM III Module Specifications

Print Specs

Displayed Oxygen Saturation Range (SpO2): 0 to 100%
Displayed Pulse Rate Range: 18 to 321 beats per minute (BPM)
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW maximum average
Infrared (using Nonin PureLight® Sensor): 910 nanometers @ 1.2 mW maximum average

SpO2 Accuracy (Arms*): 70 to 100%

Adults/Pediatrics Neonates
Motion
REUSABLE:
Finger Clip: ± 2 digits ± 3 digits
Flex: ± 3 digits ± 3 digits
Soft Sensor: ± 2 digits N/A
8000R: ± 3 digits N/A
8000Q2: ± 4 digits N/A
DISPOSABLE:
6000C Series: ± 2 digits ± 3 digits
7000 Series: ± 3 digits ± 4 digits
6500 Series: ± 2 digits N/A
Motion
REUSABLE:
Finger Clip: ± 2 digits ± 3 digits
Flex: ± 3 digits ± 4 digits
Soft Sensor: ± 3 digits N/A
Low Perfusion
All Sensors: ± 2 digits ± 3 digits

Pulse Rate Accuracy

Adults/Pediatrics Neonates
No Motion (18-300 BPM)
REUSABLE:
Finger Clip: ± 3 digits ± 3 digits
Flex: ± 3 digits ± 3 digits
Soft Sensor: ± 3 digits ± 3 digits
8000R: ± 3 digits ± 3 digits
8000Q2: ± 3 digits ± 3 digits
DISPOSABLE:
6000C Series: ± 3 digits ± 3 digits
7000 Series: ± 3 digits ± 3 digits
6500 Series: ± 3 digits N/A
Motion (40 – 240 BPM)
REUSABLE:
Finger Clip: ± 5 digits ± 5 digits
Flex: ± 5 digits ± 5 digits
Soft Sensor: ± 5 digits ± 5 digits
Low Perfusion (40-240 BPM)
All Sensors: ± 3 digits ± 3 digits

Notes:
Reusable Group:
Finger Clip Sensors: 8000AA-1, 8000AA-3, 8000AP-1, 8000AP-3
Flex Sensors: 8000J-1, 8000J-3, 8008J, 8001J
Soft Sensors: 8000SS, 8000SM, 8000SL

Disposable Group:
Flexi-Form® III (7000 Series)
Sensors: 7000A, 7000P, 7000I, 7000N
6000C Cloth Series Sensors: 6000CA, 6000CP, 6000CI, 6000CN
6500 Durafoam Series Sensors: 6500SA, 6500MA

Temperature:
Operating: 0°C to +50°C (32°F – 122°F)
• Specified operating temperature is for the module.
• Operating temperature of Sensor is not to exceed 40°C (104°F)

Storage/Transportation: -20°C to +70°C (-4°F – +158°F)

Humidity:
Operating: 10% to 90% non-condensing
Storage/Transportation: 10% to 95% non-condensing

Power Draw
29 mW typical operation @ 3.3VDC input voltage
45 mW typical operation @ 5.0VDC input voltage

Voltage Input
+3.3VDC (3.2V to 3.5V), w/50mV max. ripple
+5.0VDC ±0.25VDC, w/50mV max. ripple

I/O Signals
0 to +3.3VDC (nominal) @ +3.3VDC input voltage
0 to +5.0VDC (nominal) @ +5.0VDC input voltage

Dimensions: 1.35″ x 0.95″ x 0.235″ (34.3 x 24.1 x 6.2 mm)

Weight: 5.3g (0.19 oz.) (with shield)

Ruggedness:
Shock: IEC 60068-2-27
Ruggedness (Vibration):
Sinusoidal – IEC 60068-2-6
Random – IEC 60068-2-64

Sensors
Designed to use Nonin-branded PureLight sensors only (see Accessories)

Shielding
An RF shield is included (placed over the analog components)

* ±1 Arms represents approximately 68% of measurements.
** This information is especially useful for clinicians performing photodynamic therapy.

OEM III Module Troubleshooting

Frequently Asked Questions

Nonin
Locations

Nonin Medical Inc
(Global Headquarters)

Address
13700 1st Ave N,
Plymouth, MN 55441

Email
info@nonin.com

Toll Free
1.800.356.8874

Phone
+1.763.553.9968

Nonin Medical B.V.
(Europe)

Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
The Netherlands

Email
infointl@nonin.com

Toll Free
+31 (0)13-79 99 040

Phone
+31 (0)13-79 99 042

Interested in a career