SenSmart 8100A Series Specifications
SpO2 Accuracy (Adult/Pediatric)1, 2:
Non-motion Oxygen Saturation (Arms*) for 8100AA/8100AP:
70 – 100%: ±2
70 – 80%: ±2
80 – 90%: ±2
90 – 100%: ±2
Motion Oxygen Saturation (Arms*) for 8100AA:
70 – 100%: ±2
70 – 80%: ±2
80 – 90%: ±2
90 – 100%: ±3
Motion Oxygen Saturation (Arms*) for 8100AP:
70 – 100%: ±3
70 – 80%: ±3
80 – 90%: ±2
90 – 100%: ±2
Non-motion 95% Limits of Agreement (SpO2) for 8100AA:
70 – 100%: (3.8, 1.9)
Non-motion 95% Limits of Agreement (SpO2) for 8100AP:
70 – 100%: (3.2, 2.4)
Pulse Rate Accuracy1:
- Non-motion: 40 – 240 BPM ±3 digits (Arms*)
- Motion: 40 – 240 BPM ±5 digits (Arms*)
- Low Perfusion: 40 – 240 BPM ±3 digits (Arms*)
SpO2 Low Perfusion Accuracy1: 70 – 97% ±2 digits (Arms*)
Temperature3, 4:
- Operating: -10 °C to 40 °C (14 °F to 104 °F)
- Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)
Humidity3, 4:
- Operating: 15 to 93% non-condensing
- Storage/Transportation: Up to 93% non-condensing
Operating Altitude: 0 to 4,000 m (0 to 13,123 ft)
* ±1 Arms encompasses 68% of the population at zero bias.
- Additional accuracy and performance information can be found in the system operator’s manual.
- Accuracy specifications based on Nonin’s PureSAT® SpO2 technology and PureLight® sensor technology.
- For combined oximeter/sensor specifications, refer to the applicable oximetry system’s operator’s manual.
- Range as tested with Nonin’s PureSAT SpO2 technology.
Measurement Wavelengths and Output Power**
- Red: 660 nanometers @ 0.8 mW nominal
- Infrared: 910 nanometers @ 1.2 mW nominal
** This information is especially useful for clinicians performing photodynamic therapy.
Compliance
This product complies with ISO 10993-1.
Not made with natural rubber latex.
Warranty
1 year from the date of delivery.
The device’s expected service life is 1 year.