TruO2® OTC Model 3250 Specifications

Accuracy

  • SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits.
  • Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits.
  • Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits.
  • Clinical studies have shown that the device’s accuracy is not affected by differences in
    skin pigmentation.

Operating Conditions

  • Temperatures no lower or higher than -5°C to 40°C
    (23°F to 104°F).
  • Humidity no lower or higher than 10 to 95%.
    Battery Life
  • Approximately 2,200 uses or stored for 1 month.
  • If stored for more than 30 days, remove the batteries.

Accuracy Summary

SaO2             Measured Accuracy

70 – 100%      2.27

70 – 80%        2.85

80 – 90%        1.86

90 – 100%      1.97