Xpod® 3012LP Specifications

Displayed Oxygen Saturation Range (SpO2): 0 to 100%
Displayed Pulse Rate Range: 18 to 321 beats per minute (BPM)
Measurement Wavelengths and Output Power*:
Red: 660 nanometers @ 0.8 mW maximum average
Infrared (using Nonin PureLight® Sensor): 910 nanometers @ 1.2 mW maximum average

SpO2 Accuracy (Arms**): 70 to 100%

Adults/Pediatrics*** Neonates
Motion
REUSABLE:
8000AA Sensors: ± 2 digits N/A
8000AP Sensors: ± 2 digits N/A
800XJ Series: ± 3 digits N/A
8000SX Series: ± 2 digits N/A
8000R: ± 3 digits N/A
8000Q2: ± 3 digits N/A
DISPOSABLE:
6000 Series: ± 2 digits ± 3 digits
7000 Series: ± 2 digits ± 3 digits
6500 Series: ± 2 digits N/A
Motion
REUSABLE:
8000AA Sensors: If excessive sensor motion is occurring,
use the Model 8000J Adult Flex Sensor
N/A
8000AP Sensors: ± 5 digits N/A
800XJ Series: ± 3 digits N/A
8000SX Series: ± 3 digits N/A
Low Perfusion****
8000AX Series: ± 3 digits ± 3 digits

Pulse Rate Accuracy

Adults/Pediatrics*** Neonates
No Motion (18-300 BPM)
REUSABLE:
8000AA Sensors: ± 3 digits N/A
800XJ Series: ± 3 digits N/A
8000SX Series: ± 3 digits N/A
8000R: ± 3 digits N/A
8000Q2: ± 3 digits N/A
DISPOSABLE:
6000C Series: ± 3 digits ± 3 digits
7000 Series: ± 3 digits ± 3 digits
6500 Series: ± 2 digits N/A
Motion (40 – 240 BPM)
REUSABLE:
8000AA Sensors: If excessive sensor motion is occurring,
use the Model 8000J Adult Flex Sensor
N/A
800XJ Series: ± 5 digits N/A
8000SX Series: ± 5 digits N/A
Motion (40 – 250 BPM)
REUSABLE:
8000AP Sensors: ± 5 digits N/A
Low Perfusion (40-240 BPM)****
8000AX Series: ± 3 digits ± 3 digits

* This information is especially useful for clinicians performing photodynamic therapy.
** ±1 Arms represents approximately 68% of measurements.
*** Includes infant patients.
**** Does not apply to those sensors listed as N/A under the neonate column, 8000R and 8000Q2.

Notes: Reusable Group:
8000AX Finger Clip Sensors: 8000AA-1, 8000AA-3, 8000AP-1, 8000AP-3
800XJ Flex Sensors: 8000J-1, 8000J-3, 8008J, 8001J
8000SX Soft Sensors: 8000SS, 8000SM, 8000SL, 8000Q2, 8000R

Disposable Group:
Flexi-Form® III (7000 Series) Sensors: 7000A, 7000P, 7000I, 7000N
6000C Series Sensors: 6000CA, 6000CP, 6000CI, 6000CN
Durafoam Disposable Sensors: 6500MA, 6500SA

Temperature:
Operating: -5 to +50 °C for Xpod, 0 to +40 °C for sensor
Storage/Transportation: -40 to +70 °C

Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing

Patient Isolation: Type BF
Leakage Current: Not Applicable
Dimensions:
Cable Length: 1 meter ±10% or per customer request (2 meter maximum)
Housing: No larger than 1.1 cubic inches in volume +10%

Weight: No more than 75 grams

Power Draw (typical)*
Typical: 35mW or less with 3.3V input (non-USB configuration)
90 mW +/-10 mW for USB configurations

Input Voltage Power* mW
1.0 50
1.2 46
1.4 41
1.6 39
1.8 37
2.0 36
2.2 36
2.4 35
2.6 34
2.8 34
3.0 33
3.2 33
3.3 34
3.4 34
3.6 35
3.8 36
4.0 37
4.2 39
4.4 40
4.6 41
4.8 43
5.0 45
5.2 46
5.4 47
5.5 47

Inrush Power**: Shall not exceed an average of 275 mW within a sliding 5 mS window for any voltage input from 1.0 to 5.5 VDC.
Voltage Input: 1.0 to 5.5 VDC, w/100 mV max. ripple

* Excludes USB configurations and Range Modified (RM) Xpod LP version.
** Applies to standard Xpod LP configurations only.