6000C Specifications
Model 6000C
SpO2 Accuracy:
70 % to 100 % ±3 digits (Arms*)1 (Neonate)
70 % to 100 % ±2 digits (Arms*)2 (Adult)
SpO2 Low Perfusion Accuracy: 70 % to 100 % ±2 digits (Arms*)3
Pulse Rate Accuracy: 18 to 321 BPM ±3 digits (Arms*)3
Pulse Rate Low Perfusion Accuracy: 40 to 240 BPM ±2 digits (Arms*)3
* ±1 Arms encompasses 68 % of the population.
1Accuracy is based on 231 data pairs collected on 26 subjects, < 30 days of age and compared with co-oximetry data over the range of 75-100%. Accuracy specifications based on testing with Model 7500.
2Accuracy specifications based on testing with Models 2500/2500A and 9600.
3Additional accuracy and performance information can be found in the pulse oximeter operator’s manual.
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 3 mW nominal
Infrared: 910 nanometers @ 3 mW nominal
** This information is especially useful for clinicians.
Compliance
This product complies with ISO 10993-1.
Not made from natural rubber latex.
Models 6000CP/7000P, 6000CA/7000A, 6000CI/7000I
SpO2 Accuracy:
70 % to 100 % ±2 digits (Arms*)1, 2, 3
SpO2 Low Perfusion Accuracy: 70 % to 100 % ±2 digits (Arms*)3
Pulse Rate Accuracy: 18 to 321 BPM ±3 digits (Arms*)3
Pulse Rate Low Perfusion Accuracy: 40 to 240 BPM ±2 digits (Arms*)3
* ±1 Arms encompasses 68 % of the population.
1SpO2 accuracy was conducted during induced hypoxia study on healthy subjects
over the range of 70 % to 100 %.
2Accuracy specifications based on testing with Models 2500/2500A and 9600.
3Additional accuracy and performance information can be found in the pulse oximeter operator’s manual.
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 3 mW nominal
Infrared: 910 nanometers @ 3 mW nominal
** This information is especially useful for clinicians.
Compliance
This product complies with ISO 10993-1.
Not made from natural rubber latex.
Nonin
Locations
Nonin Medical Inc
(Global Headquarters)
Address
13700 1st Ave N,
Plymouth, MN 55441
Email
info@nonin.com
Toll Free (USA and Canada)
1.800.356.8874
Phone
1.763.553.9968
Nonin Medical B.V.
(Europe)
Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
Netherlands
Email
info@nonin.com