SenSmart 8100Q2 Specifications
SpO2 Accuracy1, 2, 3:
Range: Oxygen Saturation (Arms*)
70 – 100%: ±2
70 – 80%: ±2
80 – 90%: ±2
90 – 100%: ±2
Non-motion 95% Limits of Agreement (SpO2)
70 – 100%: (-2.4, 3.5)
Pulse Rate Accuracy1:
- Non-motion: 40 – 240 BPM ±3 digits (Arms*)
- Low Perfusion: 40 – 240 BPM ±3 digits (Arms*)
SpO2 Low Perfusion Accuracy1: 70 – 97% ±2 digits (Arms*)
Temperature4, 5:
- Operating: -10 to 40 °C (14 to 104 °F)
- Storage/Transportation: -40 to 70 °C (-40 to 158 °F)
Humidity4, 5:
- Operating: 15 to 93% non-condensing
- Storage/Transportation: Up to 93% non-condensing
Operating Altitude: 0 to 4,000 m (0 to 13,123 ft)
* ±1 Arms encompasses 68% of the population at zero bias.
- Additional accuracy and performance information can be found in the system operator’s manual.
- Accuracy testing was performed under no-motion conditions.
- Accuracy specifications based on Nonin’s PureSAT® SpO2 technology and PureLight® sensor technology.
- For combined oximeter/sensor specifications, refer to the applicable oximetry system’s operator’s manual.
- Range as tested with Nonin’s PureSAT SpO2 technology.
Measurement Wavelengths and Output Power**
- Red: 660 nanometers @ 0.8 mW nominal
- Infrared: 910 nanometers @ 1.2 mW nominal
** This information is especially useful for clinicians performing photodynamic therapy.
Compliance
This product complies with ISO 10993-1.
Not made with natural rubber latex.
Warranty
90 days from the date of delivery.
The device’s expected service life is 90 days.
Frequently Asked Questions
How to I attach the 8100Q2 sensor to the patient?
NOTES:
- For use on adult and pediatric patients weighing greater than 40 kilograms (88 pounds).
- Proper sensor placement is critical for good performance. If the sensor is not positioned properly, light may bypass the tissue and result in SpO2 inaccuracies.
1. Carefully remove the sensor from the plastic pouch and uncoil the sensor cable. Check the sensor for any sign of damage in transport. If signs of damage are found, replace the sensor.
2. Rub the ear lobe vigorously for at least 5 seconds.
3. Apply the ear clip sensor to the lobe of the ear. Ensure that the ear clip sensor is positioned so the light emitter and light detector are completely covered by the earlobe. This ensures that no ambient light bypasses the earlobe, which can cause SpO2 inaccuracies.
4. Secure the sensor cable with medical tape so the cable does not become caught on nearby equipment. Ensure that the tape securing the cable does not restrict blood flow or pull the sensor out of position.
CAUTION: As with all medical equipment, carefully route patient cables and connections to reduce the possibility of entanglement or strangulation.
5. Align the arrows on the sensor connector and the signal processor. Insert the sensor connector into the signal processor connection port.
6. Flip the clear lock over the sensor connector and click it into place.
7. Verify proper operation as described in the system operator’s manual. Verify the sensors are connected as needed for the desired system configuration and that the displayed data correctly correlates with the sensor application site.
How do I clean the 8100Q2 sensor?
Cautions:
• Clean the sensor before applying it to a new patient.
• Disconnect the sensor from the signal processor before cleaning.
• Do not sterilize, autoclave or immerse the sensor in liquid of any kind. Do not
pour or spray any liquids onto the sensor.
• Do not use caustic or abrasive cleaning agents on the sensor. Do not use any
cleaning solution other than those recommended here, as permanent damage
could result.
- To clean the sensor, wipe all patient contact surfaces with a soft cloth dampened with a 10% bleach/90% water solution (household bleach [containing less than 10% sodium hypochlorite]).
- Allow the sensor to dry thoroughly before reusing.
NOTE: To minimize cable deterioration when cleaning the cable, gently wipe away from the plug end towards the sensor end.
Nonin
Locations
Nonin Medical Inc
(Global Headquarters)
Address
13700 1st Ave N,
Plymouth, MN 55441
Email
info@nonin.com
Toll Free (USA and Canada)
1.800.356.8874
Phone
1.763.553.9968
Nonin Medical B.V.
(Europe)
Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
Netherlands
Email
info@nonin.com